AH 2500A
1 kHz Automatic
Capacitance Bridge
|
||
The World’s Most Accurate Capacitance/Loss1 Bridge |
Outstanding Features
BACKGROUND |
Unfortunately, this bridge had neither the stability nor the resolution to measure the small changes required. As a result, Andeen modified this bridge to give it much better temperature stability and to increase its resolution by more than two orders of magnitude.
During this time, Andeen taught a course where he met Carl Hagerling, a student in his class. They appreciated each other’s technical abilities and quickly struck up a close friendship.
Driven by the needs of his colleagues for better capacitance bridge technology for their research, Andeen continued throughout the 1970’s to hone his expertise in this area by building numerous prototype bridges. By 1980, he was building a prototype of a fully automated, high performance, ratio-transformer based capacitance bridge. It was apparent that there would be a demand for a production version of such a bridge and that Andeen would need help meeting that demand.
Fortunately, in this same year, Hagerling graduated with his Ph.D. in physics. He and Andeen soon committed to form a company to manufacture a production version of Andeen’s high performance bridge.
Andeen-Hagerling, Inc. was incorporated in 1982. The company name is composed of the names of the founders to underscore what they knew from their past experience would be a long, stable and productive relationship.
Andeen-Hagerling is fortunate to be privately owned, and thus able to make and follow through with long-term development plans. The company’s dedicated employees take pride in Andeen-Hagerling’s exceptionally high standards and uncompromising view of quality, reliability, and customer satisfaction. The company is proud to be able to support its instruments which are in use in national standards laboratories and industrial, government, university and military facilities throughout the world.
Andeen-Hagerling is headquartered in the city of Solon, a suburb of Cleveland, Ohio, U.S.A.
BASIC DESIGN
The Model 2500A measures capacitance and loss in medium- and high-impedance ranges, and thus allows using three terminal rather than five terminal connections to the unknown. Its unmatched precision is the result of a uniquely designed ratio transformer which is the culmination of 15 years of custom bridge design and manufacture. Equally important is the unique temperature-controlled, fused silica capacitance standard which allows extremely high measurement stability and immunity to mechanical shock. These elements combine to form a true bridge operating at one kilohertz to give capacitance results which are independent of the exact test frequency.
MEASUREMENT FEATURES
Measurement Initiation
A single measurement is initiated by a front panel keystroke, an external trigger pulse, a single character from the RS-232 or IEEE-488 ports, or a Group Execute Trigger from the IEEE-488 bus. Measurements can be taken continuously with a selectable delay time between the end of one reading and the start of the next. This delay time can range from zero to many hours in 0.01 second increments.
Units
Capacitance units are picofarads. Loss units are selectable among nanosiemens, dissipation factor, series resistance in kilohms, parallel resistance in gigohms or magnitude of the loss vector in jpF — the choice being indicated by front panel LED’s.
Display Results
Results are displayed on large, variable brightness front panel LED displays to as many as eight digits. Results are sent to remote devices with as many as nine digits.
Deviation Measurements
Results may be provided in the form of a difference or offset from a reference value for capacitance or loss or both. The loss may be expressed using any of the 2500A’s loss units. The reference value can be the result of a previous measurement or a user-provided value.
Zero Correction
Stray capacitance and loss (typically associated with a test fixture) may occur in parallel with the capacitance and loss that is to be measured. The stray values can be obtained from the result of a previous measurement or from a user-provided value and used to correct the reported results. The stray loss is corrected for as if it is in parallel with the loss that is intended to be measured. This occurs no matter what loss units are being used. This is more involved than a deviation measurement which would just do a simple subtraction. (The 2500A itself has no significant zero offset.)
DC Bias
A connector is provided to which an external DC bias voltage may be applied. The 2500A can switch this voltage to the sample through user selectable resistors located within the instrument.
Test Voltage
The maximum test voltage applied to the sample is selectable from 0.5 mV to 15 V r.m.s. The actual voltage applied by the 2500A may be much smaller than the selected maximum.
Speed versus Resolution
Available resolution is determined primarily by the amount of time spent averaging out noise. The trade-off between resolution and measurement speed is selectable in factors of two from less than 40 milliseconds to 20 minutes.
Commutation
This selectable feature causes the test signal to be reversed periodically so as to improve rejection of external periodic signals, particularly those that are power-line related.
Tracking
In the case of changing or rapidly drifting samples, long averaging times have little meaning. Thus a tracking feature is provided which allows samples to be rapidly followed at a rate of about 25 measurements per second with reduced resolution. Tracking occurs automatically when this feature is enabled and the value of the sample is changing.
Bridge Balancing Time
Measurement time on a previously unmeasured sample is less than half a second. However, the measurements following the first can be made in less than 40 milliseconds if the averaging time is short.
Standards Oven
The oven (and hence the entire bridge) is normally stable within only 15 minutes after power-on. A blinking front panel LED indicates when the oven has not stabilized or when the ambient temperature is too extreme for stabilization.
Cable Length Correction
The three terminal connection method used by the 2500A usually makes the errors caused by the pair of cables that connect the instrument to the unknown capacitance so small that they can be ignored. However, cable inductance can affect the accuracy of capacitance measurements made at the high end of the 2500A’s range. Similarly, cable resistance can affect the accuracy of loss measurements made at the high end of the 2500A’s loss range. In these situations, the resistance and inductance per meter of cable pair and the length of the cable pair can be entered into the instrument. The 2500A will then automatically correct for these cable errors.
Measurement Errors
Measurement troubles are easily pinpointed by one of over a dozen English language error messages (or, optionally, error codes). Additionally, many other command and status messages are reported.
Calibration
A unique calibration technique allows internal precision components to be compared against internal temperature-controlled standards with the appropriate corrections being made by a microprocessor. The 2500A also provides for calibration against external standards. To prevent unauthorized calibrations, a passcode (which only the manager of the instrument can change) must be entered before any calibration can be performed.
Self Tests
Power-on or user initiated self tests check the microprocessor area, transformer ratio-arm switches, D/A switches and A/D converter. Special circuitry allows numerous internal self-consistency checks.
SYSTEM INTERFACES
RS-232
An RS-232 standard serial interface is included with each instrument to allow simple connections to a computer, modem, printing terminal or video terminal. These devices can take control of the instrument interactively or can merely log the measured data passively.
IEEE-488
An IEEE-488 (1978) standard interface is included with each instrument to allow connection to an instrument bus. A full IEEE-488 implementation is provided including serial poll and selectable extended talker/listener addressing. The 2500A can be run with a bus controller or can operate in “talk only” mode to send data to a passive printer or data logger. Front panel “remote”, “talk”, and “listen” indicators are provided.
Setup of IEEE-488 and RS-232
IEEE-488 bus address and RS-232 baud rate, parity, stop bits, and fill characters are all entered from the front panel keypad and can be permanently stored from the keypad as well.
Friendly Commands
Both remote device interfaces use the same English language commands that are found on the front panel. Commands can be abbreviated by supplying only enough letters to uniquely identify the desired command.
Data Formats
Measurement results consist of any combination of four fields: error message, capacitance, loss, and voltage. The number of decimal places and the width of the capacitance and loss fields are independently selectable. Field and unit labels are optional. Numeric results can be reported in floating point, scientific or engineering notations.
THE 2500A HAS MANY POSSIBLE USES BEYOND CALIBRATION
The reaction of many technical persons upon first learning of the Model 2500A is: “That’s a very impressive instrument, but we don’t see a need for such precision in our work. Furthermore, such measurements must be more difficult to make.” Until the introduction of the 2500A, this attitude toward high precision capacitance measurements was justified. Previously, the only commercially available instruments were manually operated, required a skilled operator to spend several minutes balancing the bridge, were prone to reliability problems due to the large number of open switching contacts used, and were still far less stable than the 2500A. It is not surprising that these bridges have not seen significant use outside of calibration or research laboratories.
Today, the incredible ease with which high precision capacitance and loss measurements can be made with the 2500A requires a reassessment of previous attitudes. The 2500A allows totally automated operation with no human intervention. Its ability to maintain its precision over a wide temperature range and its immunity to mechanical shock make it ideally suited for factory-floor or portable field use. Andeen-Hagerling is so confident of its ability to perform reliably, that it is provided with a three year warranty.
To apply the 2500A to a productive task requires obtaining a suitable sensor. This is where the possibilities become exciting, because capacitive sensors are theoretically the most precise of all electrical sensors. The reasons are:
EXAMPLES OF APPLICATIONS OF THE MODEL 2500A
SPECIFICATIONS
Notation: | The specifications are grouped according to whether the unknown is modeled as a resistor in parallel with a capacitor or in series with it. | |
Parallel: | “C” | is the value of the unknown (parallel) capacitance in picofarads (pF = 10-12 F). Also used are attofarads (aF = 10-6 pF) and microfarads (µF = 106 pF). |
“G” | is the value of the unknown loss expressed as a conductance in nanosiemens (nS = 10-9 S). | |
“D” | is the value of the unknown loss expressed as a dissipation factor (tan delta). D has no units. | |
“RP” | is the value of the unknown loss expressed as a parallel resistance in gigohms (G ohm = 109 ohm). | |
Series: | “CS” | is the value of the unknown series capacitance in picofarads (pF = 10-12 F). |
“RS” | is the value of the unknown loss expressed as a series resistance in kilohms (k ohm = 103 ohm). | |
Misc: | “tb” | is the minimum selectable time between consecutive measurements in seconds. |
“V” | is the AC test signal voltage in volts applied across the unknown. Its upper limit may be selected by the user to have any value listed in the AC Test Signal Voltages table. | |
“ppm” | means Parts Per Million. |
General:
The expressions below for accuracy, linearity, stability, resolution, and temperature coefficient give absolute rather than statistical uncertainties. Absolute uncertainties are the most conservative of those in common use. Andeen-Hagerling guarantees repair within the warranty period of any Model 2500A whose measured errors repeatedly exceed these uncertainties. The expressions may be evaluated for particular values of capacitance (C or CS), loss (G, D, RP or RS), test voltage (V), and measurement time (t). Only the resolution expressions contain the measurement time. However, the other uncertainty expressions assume that the measurement time has been set to be long enough so that these other uncertainties are not limited by the resolution specification. In other words, specifications such as accuracy may be limited by the resolution rather than the accuracy expression if the measurement time is set too short.
Most of the uncertainty expressions can be evaluated by direct substitution of the values of capacitance, loss and voltage as if they were read directly from the 2500A. The instrument reports these values in the units given in the notation section above. Some expressions also require the dissipation factor, D, which, if it is not directly available, can be calculated using one of the following relations:
Given units of | D: | use | G = 2 pi CD |
RP: | G = 1/RP | ||
RS: | G = 2 pi CSD/(1+D2) | ||
CS: | C = CS/(1+D2) |
A comprehensive set of contour plots of all of the uncertainty expressions is available from Andeen-Hagerling upon request. Accuracy, stability, linearity and resolution specifications assume a recent internal calibration at the operating temperature.
Range:2
Parallel: | C: | -0.0012/|D| µF to +0.012/|D| µF for D >= 0.01 |
-0.12 µF to +1.2 µF for -0.001 <= D < 0.01 | ||
-0.12 µF to +0.0012/|D| µF for -0.1 <= D < -0.001 | ||
-0.012/|D| µF to +0.0012/|D| µF for D < -0.1 | ||
The capacitance range is also shown graphically in Figure 1. | ||
G: | -6000 nS to +60 000 nS | |
D: | See Figure 1. | |
RP: | -1.7x10-4 G ohm to -1.7x106 G ohm | |
and +1.7x10-5 G ohm to +1.7x106 G ohm | ||
Series: | CS: | See Figure 2. |
RS: | See Figure 2. |
[2] The ranges of all measurable variables except RP cover a region defined by negative numbers for the lower limit and positive numbers for the upper limit. This is due to the Model 2500A’s ability to measure both positive and negative values of capacitance and loss. Other instruments which only measure positive values have ranges which cover a region defined by small positive numbers for the lower limits to large positive numbers for the upper limits. For the 2500A, the small numbers which correspond to the lower limits of other instruments are given by the 2500A’s resolution specifications in absolute units.
Front Panel Display Limitations:
(The front panel display may further limit the range and resolution of the capacitance and loss.) | ||
Capacitance: | 0.1 aF is best display resolution for C and CS. | |
Loss: | G: | 10-7 nS is best conductance display resolution. |
D: | 10-7 is best dissipation display resolution. | |
RS: | 10-7 k ohm is best series resistance display resolution. | |
RP: | 10-7 G ohm is best parallel resistance display resolution. |
Remote Device Reporting Limitations:
Capacitance: | 0.01 aF is best resolution for C and CS. | |
Loss: | G: | 10-8 nS is best conductance resolution. |
D: | 10-8 is best dissipation resolution. | |
RS: | 10-7 k ohm is best series resistance resolution. | |
RP: | 10-8 G ohm is best parallel resistance resolution. |
Figure 1. The measurable values of C and D are enclosed within the shaded area. |
Figure 2. The values of CS
and RS are measurable in the six shaded regions. In five of
these regions, one or both of the measured values are too large to report
on the 2500A’s display. In three of these five regions, one or both values
are also too large to send to any remote devices. The table below shows
what can be reported in each region. A “Display” entry means that the result
can be shown on the instrument’s display. A “Remotes” entry means that
the result can be reported to an RS-232 or IEEE-488 device.
|
Measurement Time:
Parallel: | C: | {0.15C + 50DC + [7.5(1+nC) + nVC]/V}x10-6 pF |
G: | {50G + C + 5x10-5C2 + [50(1+nC) + 6nVC]/V}x10-6 nS | |
D: | {8x10-6C + (1+D2)1/2[0.15 + 50D + (7.5(1+nC)/C + nV)/V]}x10-6 | |
RP: | RP{50 + RP[C + 5x10-5C2 + (50(1+nC) + 6nVC)/V]}x10-6 G ohm | |
Series: | CS: | {0.15 + 50D + [7.5(1+nC)(1+D2)/CS + nV]/V}CSx10-6 pF |
RS: | {1.3 + 50RS + [0.15 + [7.5(1+nC)(1+D2)/CS + nV]/V]RS/D}x10-6 k ohm | |
where nC = 1.4t -1/2 and nV = 0.01(RS+10)1/2(1+D2)1/2t -1/2. | ||
t = tb except when tb = 0.05 in which case t = tb/4. | ||
The series resistance RS may be calculated for the parallel expressions using Rs = 1.6x105D/C(1+D2). |
Resolution in ppm:4
Parallel: | C: | 0.15 + 50D + [7.5(1+nC)/C + nV]/V |
G: | 50 + {C + 5x10-5C2 + [50(1+nC) + 6nVC]/V}/G | |
D: | {8x10-6C + (1+D2)1/2[0.15 + 50D + (7.5(1+nC)/C + nV)/V]}/D | |
RP: | 50 + RP{C + 5x10-5C2 + [50(1+nC) + 6nVC]/V} | |
Series: | CS: | 0.15 + 50D + [7.5(1+nC)(1+D2)/CS + nV]/V |
RS: | 1.3/RS + 50 + {0.15 + [7.5(1+nC)(1+D2)/CS + nV]/V}/D |
[4] Resolution is the smallest repeatable difference in readings that is guaranteed to be measurable at every capacitance or loss value. Useful resolution is typically a factor of ten better.
Non-linearity in ppm:
Parallel: | C: | ±{0.15 + 50D + 7.5/CV + 15x10-6C} |
G: | ±{50 + [C + 5x10-5C2 + 50/V]/G} | |
D: | ±{8x10-6C + (1+D2)1/2[0.15 + 50D + 7.5/CV]}/D | |
RP: | ±{50 + RP[C + 5x10-5C2 + 50/V]} | |
Series: | CS: | ±{0.15 + 50D + 7.5(1+D2)/CSV + 15x10-6CS/(1+D2)} |
RS: | ±{1.3/RS + 50 + [0.15 + 7.5(1+D2)/CSV]/D} |
Non-linearity is the deviation from a best fit straight line through a plot of the measured quantity versus the actual quantity. The test signal voltage is assumed to be constant.
Accuracy in ppm following calibration:
Parallel: | C: | ±{5 + 200D + (0.2 + 7.5/C)/V} |
G: | ±{200 + [13C + 0.002C2 + (45 + 1.2C)/V]/G} | |
D: | ±{2 + 3x10-4C + (1+D2)1/2[200D + (0.2 + 7.5/C)/V]}/D | |
RP: | ±{200 + [13C + 0.002C2 + (45 + 1.2C)/V]RP} | |
Series: | CS: | ±{5 + 200D + [0.2 + 7.5(1+D2)/CS]/V} |
RS: | ±{200 + 50/RS + [2 + (0.2 + 7.5(1+D2)/CS)/V]/D} |
The length of the cables connecting the 2500A to the capacitance being measured has a negligible effect on the accuracy for small capacitances. This assumes that the coaxial shield on these cables has 100% coverage. If uncorrected by the CABLE command, cables similar to RG-58 will increase the capacitance readings by about 40 ppm per meter of cable pair and per µF of capacitance being measured.
The accuracy Y years following calibration may be calculated from the expression A + YS where A is the desired accuracy expression from above and S is the corresponding stability per year below.
Stability in ppm per year:
Parallel: | C: | ±{1 + 30D + (0.01 + 2.5/C)/V} |
G: | ±{30 + [2C + 3x10-4C2 + (15 + 0.06C)/V]/G} | |
D: | ±{0.3 + 5x10-5C + (1+D2)1/2[30D + (0.01 + 2.5/C)/V]}/D | |
RP: | ±{30 + [2C + 3x10-4C2 + (15 + 0.06C)/V]RP} | |
Series: | CS: | ±{1 + 30D + (0.01 + 2.5(1+D2)/CS)/V} |
RS: | ±{30 + 8/RS + [0.3 + (0.01 + 2.5(1+D2)/CS)/V]/D} |
Temperature coefficient relative to change
in ambient temperature in ppm per °C:
Parallel: | C: | ±{0.025 + 30D + 0.002/V + 15/(0.15 + CV)} |
G: | ±{30 + [0.2 + 2x10-5C + 0.012/V]C/G + 100/(1 + GV)} | |
D: | ±{[0.03 + 3x10-6C + (1+D2)1/2(30D + 0.002/V)]/D + 15/(0.15 + CV) + 15/(0.15 + CDV)} | |
RP: | ±{30 + [0.2 + 2x10-5C + 0.012/V]CRP + 100/(1 + V/RP)} | |
Series: | CS: | ±{0.025 + 30D + 0.002/V + 100/[1 + 6CSV/(1+D2)]} |
RS: | ±{30 + 0.5/RS + (0.03 + 0.002/V)/D + 30/[0.15(1+D2) + CSV] + 100/[1 + 106VD2/(1 + D2)RS]} |
Sensitivity to changes in power line voltage:
Capacitance: | ±0.002 ppm per 1% change in line voltage |
Loss: | Not measurable |
DC Bias: Up to ±100 volts may be applied to the unknown through the external DC bias input.
Frequency: 1.0000 ±0.005% kilohertz
Operating temperature range: 0° to 45°C
Storage temperature range: -40° to +75°C
Humidity: 0 to 85% relative humidity, non-condensing
AC Test Signal
Voltages: Any voltage listed below may be selected. The capacitance
and loss ranges measurable at the selected voltage are shown. The 2500A
will automatically use the lesser of the user’s selected voltage or the
highest voltage listed in the table which provides sufficient range to
be able to measure the capacitance and loss of the unknown. The voltages
listed have tolerances of ±5%.
Voltage: | Range of C: | Range of G: | |||||||||
15.0 | V | -8 | to | +80 | pF | -0.4 | to | +4 | nS | ||
7.50 | V | -16 | to | +160 | pF | -0.8 | to | +8 | nS | ||
3.75 | V | -16 | to | +160 | pF | -0.8 | to | +8 | nS | ||
3.00 | V | -40 | to | +400 | pF | -2 | to | +20 | nS | ||
1.50 | V | -80 | to | +800 | pF | -4 | to | +40 | nS | ||
0.750 | V | -160 | to | +1600 | pF | -8 | to | +80 | nS | ||
0.375 | V | -160 | to | +1600 | pF | -8 | to | +80 | nS | ||
0.250 | V | -480 | to | +4800 | pF | -24 | to | +240 | nS | ||
0.125 | V | -480 | to | +4800 | pF | -24 | to | +240 | nS | ||
0.100 | V | -1200 | to | +12 000 | pF | -60 | to | +600 | nS | ||
0.050 | V | -1200 | to | +12 000 | pF | -60 | to | +600 | nS | ||
0.030 | V | -4000 | to | +40 000 | pF | -200 | to | +2000 | nS | ||
0.015 | V | -4000 | to | +40 000 | pF | -200 | to | +2000 | nS | ||
0.010 | V | -12 000 | to | +120 000 | pF | -600 | to | +6000 | nS | ||
0.0050 | V | -12 000 | to | +120 000 | pF | -600 | to | +6000 | nS | ||
0.0030 | V | -40 000 | to | +400 000 | pF | -2000 | to | +20 000 | nS | ||
0.0015 | V | -40 000 | to | +400 000 | pF | -2000 | to | +20 000 | nS | ||
0.0010 | V | -120 000 | to | +1200 000 | pF | -6000 | to | +60 000 | nS | ||
0.0005 | V | -120 000 | to | +1200 000 | pF | -6000 | to | +60 000 | nS |
Power requirements: 25 watts
Power frequency: 48 to 440 Hz
Power voltage ranges: 85 to 115, 102 to 138, 187 to 253 and 204 to 276 volts rms
Packaging: The instrument is 3.5 inches (8.9 cm) high and 15 inches (38.1 cm) deep behind the front panel. Hardware for rack mounting and a bail for bench top use are provided.
Weight: 18 pounds (8.2 kg)
Safety: Designed in accordance with UL1244, IEC348 and BS4743.
Radiated emissions: Designed and tested to meet FCC and VDE class A requirements.
Patents: The Model 2500A is protected by U.S. Patent No. 4 772 844. Foreign patents are pending.
Warranty: The Model 2500A is covered by a three year warranty. Forward and return shipping is covered during the first three months of the warranty.
Note: Specifications are subject to change without notice.
OPTIONS
The 2500A may be ordered with Option E which enhances the precision of the bridge. These enhanced specifications are listed below. All other specifications remain unchanged. Notes related to the specifications below may be found in the specifications above for the standard version of the bridge.
Resolution in absolute units:
Parallel: | C: | {0.07C + 20DC + (7.5(1+nC) + nVC)/V}x10-6 pF |
G: | {20G + 0.4C + 5x10-5C2 + (50(1+nC) + 6nVC)/V}x10-6 nS | |
D: | {8x10-6C + (1+D2)1/2[0.07 + 20D + (7.5(1+nC)/C + nV)/V]}x10-6 | |
RP: | RP{20 + RP[0.4C + 5x10-5C2 + (50(1+nC) + 6nVC)/V]}x10-6 G ohm | |
Series: | CS: | {0.07 + 20D + [7.5(1+nC)(1+D2)/CS + nV]/V}CSx10-6 pF |
RS: | {1.3 + 20RS + {0.07 + [7.5(1+nC)(1+D2)/CS + nV]/V}RS/D}x10-6 k ohm |
Resolution in ppm:
Parallel: | C: | 0.07 + 20D + [7.5(1+nC) /C + nV]/V |
G: | 20 + {0.4C + 5x10-5C2 + [50(1+nC) + 6nVC]/V}/G | |
D: | {8x10-6C + (1+D2)1/2[0.07 + 20D + (7.5(1+nC)/C + nV)/V]}/D | |
RP: | 20 + RP{0.4C + 5x10-5C2 + [50(1+nC) + 6nVC]/V} | |
Series: | CS: | 0.07 + 20D + [7.5(1+nC)(1+D2)/CS + nV]/V |
RS: | 1.3/RS + 20 + {0.07 + [7.5(1+nC)(1+D2)/CS + nV]/V}/D |
Non-linearity in ppm:
Parallel: | C: | ±{0.07 + 20D + 7.5/CV + 5x10-6C} |
G: | ±{20 + [0.4C + 5x10-5C2 + 50/V]/G} | |
D: | ±{8x10-6C + (1+D2)1/2[0.07 + 20D + 7.5/CV]}/D | |
RP: | ±{20 + RP[0.4C + 5x10-5C2 + 50/V]} | |
Series: | CS: | ±{0.07 + 20D + 7.5(1+D2)/CSV + 5x10-6CS/(1+D2)} |
RS: | ±{1.3/RS + 20 + [0.07 + 7.5(1+D2)/CSV]/D} |
Accuracy in ppm following calibration:
Parallel: | C: | ±{3 + 100D + (0.01 + 7.5/C)/V} |
G: | ±{100 + [13C + 0.002C2 + (45 + 0.06C)/V]/G} | |
D: | ±{2 + 3x10-4C + (1+D2)1/2[100D + (0.01 + 7.5/C)/V]}/D | |
RP: | ±{100 + [13C + 0.002C2 + (45 + 0.06C)/V]RP} | |
Series: | CS: | ±{3 + 100D + [0.01 + 7.5(1+D2)/CS]/V} |
RS: | ±{100 + 50/RS + [2 + (0.01 + 7.5(1+D2)/CS)/V]/D} |
Stability in ppm per year:
Parallel: | C: | ±{0.5 + 20D + (0.003 + 2.5/C)/V} |
G: | ±{20 + [2C + 2x10-4C2 + (15 + 0.02C)/V]/G} | |
D: | ±{0.3 + 3x10-5C + (1+D2)1/2[20D + (0.003 + 2.5/C)/V]}/D | |
RP: | ±{20 + [2C + 2x10-4C2 + (15 + 0.02C)/V]RP} | |
Series: | CS: | ±{0.5 + 20D + [0.003 + 2.5(1+D2)/CS]/V} |
RS: | ±{20 + 5/RS + [0.3 + [0.003 + 2.5(1+D2)/CS]/V]/D} |
Temperature coefficient relative to change
in ambient temperature in ppm per °C:
Parallel: | C: | ±{0.008 + 10D + 0.001/V + 2/(0.15 + CV)} |
G: | ±{10 + [0.2 + 2x10-5C + 0.006/V]C/G + 15/(1 + GV)} | |
D: | ±{[0.03 + 3x10-6C + (1+D2)1/2(10D + 0.001/V)]/D + 2/(0.15 + CV) + 2/(0.15 + CDV)} | |
RP: | ±{10 + [0.2 + 2x10-5C + 0.006/V]CRP + 15/(1 + V/RP)} | |
Series: | CS: | ±{0.008 + 10D + 0.001/V + 15/[1 + 6CSV/(1+D2)]} |
RS: | ±{10 + 0.5/RS + (0.03 + 0.001/V)/D + 4/[0.15(1+D2) + CSV] + 15/[1 + 106VD2/(1 + D2)RS]} |
Ordering Information: | Part No. |
Model 2500A Capacitance Bridge | 2500A |
Enhanced performance option for Model 2500A | Option E |
Consult factory or your sales representative for price and availability of Option E for bridges that were not originally purchased with this option. |
For questions regarding the AH 2500A, possible applications, the location of your nearest sales representative, or ordering information: | |